PART THREE - OPERATION
SECTION 6 - MAINTENANCE MENU
Model P63 pH Analyzer
Rev. 7-1201
84
6.7 RESET OVERFD
TIMER (reset relay
overfeed timers)
☞
this screen, the relays and TTL outputs automati-
cally return to their previous states.
1.
After selecting “Reset Overfd Timer” from the MAINTE-
NANCE menu, use the following submenu screen to
manually reset all “timed out” relay overfeed timers at
once:
This screen shows the present status for each relay.
The status possibilities are:
•
DISABLED:
Indicates relay overfeed timer has a
preset configuration value of “0.0” minutes, or the
relay has been set to function as an ALARM relay.
•
XXX.X min:
Indicates the in-progress “count down”
(in minutes) of an active relay overfeed timer from its
preset setting to zero.
•
TIMED OUT:
Indicates overfeed timer has com-
pleted counting down and needs to be reset after
investigating the cause for this “timed out” condition.
2.
With the “RESET ALL TIMEOUTS” line highlighted,
press
ENTER key
to simultaneously reset all relay
overfeed timers with a “TIMED OUT” status. Resetting
the relay overfeed timers:
•
Returns relay to its normal operation.
•
Clears “Overfeed (A, B, C) Timeout” Fail message
from message list, and NAMUR TTL Output C (if
used) responds to this condition by turning off un-
less other fail or warn conditions are pending.
•
Adds an “All Timeouts Reset” entry to the logbook.
•
Changes Analog Output 2 (only when set to repre-
sent a dedicated alarm signal) to the selected non-
alarm state value (4 mA or 20 mA).
NOTE:
See PART THREE, Section 8 for more operating
details for overfeed timers.