PART THREE - OPERATION
SECTION 6 - MAINTENANCE MENU
Rev. 7-1201
Model P63 pH Analyzer
83
6.5 MANUAL ANALOG
OUT (set/jog analog
output mA values)
☞
6.6 MANUAL RELAY/TTL
(manually activate
relays/TTL outputs)
☞
After selecting “Manual Analog Out” from the MAINTE-
NANCE menu, use the following submenu screen to manu-
ally set/jog both analog output mA values:
You can “SET” or “JOG” the analog output values.
•
SET:
Enables you to adjust the 4.00-20.00 mA value for
each output set. (0-5 VDC/0-1 mA values are also
automatically adjusted to values corresponding to the
entered 4.00-20.00 mA value.)
•
JOG:
Enables you to jog the 4.00-20.00 mA value for
each output set up or down by respectively holding
down the
×
×
key or
Ø
Ø
key
. (0-5 VDC/0-1 mA values are
also automatically adjusted to values corresponding to
the adjusted 4.00-20.00 mA value.)
NOTE:
Using this screen to set or jog the outputs overrides
and clears any “HOLD” or “XFER” setting. When
exiting this screen, both analog Outputs 1 and 2
automatically return to “ACTIVE” operation.
After selecting “Manual Relay/TTL” from the MAINTE-
NANCE menu, use the following submenu screen to manu-
ally operate the relays and TTL outputs:
By manually operating the relays and TTL outputs, you can
activate the devices connected to them.
NOTE:
Using this screen to select states for the relay and
TTL outputs temporarily overrides any “XFER” state
setting. This also will interact with relay overfeed
timer operation (if counting). Refer to PART THREE,
Section 8.5 for complete details. When exiting