![GLI P63 Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/gli/p63/p63_operating-instructions-manual_2231741062.webp)
PART THREE - OPERATION
SECTION 5 - CONFIGURE MENU
Rev. 7-1201
Model P63 pH Analyzer
61
•
Aux Wash 2 Delay:
Sets the time period (0-999 seconds)
between the “Aux Wash 2” step and re-inserting the elec-
trode. If “0” is entered, this step will be bypassed and the
electrode will be immediately re-inserted. (The monitored
TTL inputs determine when the electrode is fully inserted.
If insertion takes longer than 25 seconds, the cycle aborts
and an error message screen appears. After the electrode
is inserted, the entered time for the “Hold Release” step
counts down. When it expires, analog Outputs 1 and 2
return to their active states and the cycle ends.)
•
Cycle Interval: Only for use when “YES” is selected
for the third function step “Cycle:” to enable repetitive
cycling.
The “Cycle Interval:” function step sets the time
period (0-999 seconds) between the end of one cycle and
the beginning of the next cycle.
DEFINE CAL CYCLE Steps
for GLI Cal-Clean™ System
A “Cal” cycle is similar to a “Wash” cycle except that it also
includes function steps pertaining to calibration. Use the
following submenu screen to define and enter a “Cal” cycle:
Many of the “Cal” steps have the same names and functions
as the “Wash” cycle steps. However, they do operate inde-
pendently of each other. This provides greater flexibility
such as being able to wash the electrode anytime instead of
only being able to wash it during a “Cal” cycle. Also, a
stronger wash solution may be desired during a “Cal” cycle.
Each available “Cal” cycle function step is listed below in
the sequence shown on the “DEFINE CAL CYCLE” sub-
menu.
•
Abort Pos:
See “Wash” cycle for description.
•
Cal Timeout:
Sets how long (0-999 seconds) you want
the analyzer to wait for a successful calibration to be
completed. If the entered time expires before completing a
successful calibration, the “Cal” cycle will abort and a
message screen will appear.