![GLI P63 Скачать руководство пользователя страница 78](http://html.mh-extra.com/html/gli/p63/p63_operating-instructions-manual_2231741078.webp)
PART THREE - OPERATION
SECTION 5 - CONFIGURE MENU
Rev. 7-1201
Model P63 pH Analyzer
77
5.11 Setting MEASURE
SCREEN Resolution
5.12 Setting
Analyzer Clock
(SET TIME/DATE)
☞
After the set time elapses, the analyzer reminds you to cali-
brate by displaying a “WARN” alert, and creates diagnostic
message # 80 (cal date reached). This message can be
displayed by accessing the “Messages” submenu in the
DIAGNOSTICS menu.
After selecting “Meas Screen Setup” from the CONFIGURE
menu, use the following submenu to set display resolution
(one or two places after decimal point) for the MEASURE
screen:
Regardless of this setting, all setup screens always show
values with a two-place display resolution (XX.XX).
After selecting “Set Time/Date” from the CONFIGURE
menu, use the following submenu to set the present time
and date for the analyzer real-time clock (RTC):
NOTE:
The “12-Hour” selection in the “Mode” field provides
an am/pm time format. The “24-Hour” selection
provides a military time format.