PART THREE - OPERATION
SECTION 6 - MAINTENANCE MENU
Rev. 7-1201
Model P63 pH Analyzer
85
6.8 SIMULATE pH/TEMP
Values
☞
After selecting “Simulate pH/Temp” from the MAINTE-
NANCE menu, use the following submenu screen to simu-
late pH and temperature values:
Simulating pH and
°
C values is useful when setting up and
verifying control or alarm schemes. Example: Suppose
you’ve set an alarm relay to sound a horn when the meas-
ured pH exceeds 12 pH. By using a simulated pH value, you
can verify this condition without actually upsetting the proc-
ess value. A “SIM” annunciator appears on the MEASURE
screen to remind you of a simulated condition.
NOTE:
Using this screen to manually simulate values may
interact with relay overfeed timer operation (if
counting). Refer to PART THREE, Section 8.5 for
complete details.