PART THREE - OPERATION
SECTION 5 - CONFIGURE MENU
Model P63 pH Analyzer
Rev. 7-1201
66
☞
For GLI
Air/Water Blast
Cleaning System
NOTE:
You must clear the alarm, and use the “Manual
Insert” function (in the “Sensor Maintenance” sub-
menu listed in the MAINTENANCE menu) to insert
the electrode into the process and clear the output
hold.
Similarly, if the electrode is not in the position that it is ex-
pected to be in during a “Wash” or “Cal” cycle, or when
manually retracting the sensor, the cycle aborts and an er-
ror message screen appears.
The analyzer software provides a wash-only function that
controls an optional GLI Air/Water Blast Cleaning System to
automatically wash a GLI Differential Technique pH sensor
with user-supplied air or water. You can establish when the
wash (air or water blast) occurs and the duration of the
wash.
When you enable wash system operation, Relay C is used
to control the GLI Air/Water Blast Cleaning System attached
to the sensor. During the wash cycle, analog Outputs 1 and
2 and Relays A and B are automatically held at their pres-
ent states to prevent upset of control or recording
equipment that may be connected to them. After the com-
pleted wash cycle, the analog outputs and relays are
released and returned to their configured “TRANSFER
CONDITION” states unless you want to continue holding
them for up to 999 seconds.
To enable the wash system feature, press
CONFIG key
to
display the CONFIGURE main menu, select the “Wash
System” line, press
ENTER key
to display the following
submenu screen, and select “YES.”