PART TWO - INSTALLATION
SECTION 3 - ELECTRICAL CONNECTIONS
Model P63 pH Analyzer
Rev. 7-1201
30
Solid State DC
Relays (optional)
☞
Three sets of SPST solid state DC relay outputs (Relays A,
B and C) are provided at Terminals 1 through 8 on TB4 --
except for unused Terminals 3 and 6.
The relay outputs
are not powered.
You must supply DC power to the control
or alarm devices with these relay contacts. Refer to Figure
2-10 for a general wiring arrangement. Always check control
wiring to insure that the DC power source will not be
shorted by the relay switching action, and that wiring con-
forms to local codes.
NOTE:
These solid state DC relay outputs require 3-60
VDC power.
FIGURE 2-10
Connecting Control/Alarm Device(s) to
Solid State DC Relays