![GLI P63 Скачать руководство пользователя страница 60](http://html.mh-extra.com/html/gli/p63/p63_operating-instructions-manual_2231741060.webp)
PART THREE - OPERATION
SECTION 5 - CONFIGURE MENU
Rev. 7-1201
Model P63 pH Analyzer
59
☞
The “Schedule Wash/Cal” submenu is used to schedule
“Wash” and “Cal” cycles on a time basis.
NOTE:
Use the “Remote System Cntrl” (control) submenu
only when you intend to use a TTL input signal to
remote start a “Wash” or “Cal” cycle.
DEFINE WASH CYCLE Steps
for GLI Cal-Clean™ System
Use the following submenu screen to define and enter a
“Wash” cycle:
Each available “Wash” cycle function step is listed below in
the sequence shown on the “DEFINE WASH CYCLE” sub-
menu.
•
Abort Pos:
Sets the default position (RETRACTED or
INSERTED) that you want the electrode to be in if a cycle
is aborted. With “Abort Pos:” highlighted, use
×
×
or
Ø
Ø
key
to select the choice.
•
Hold Pos:
Sets the position (RETRACTED or INSERT-
ED) that you want the electrode to be held in during the
“Hold Release:” function step. With “Hold Pos:” high-
lighted, use
×
×
or
Ø
Ø
key
to select the choice.
•
Cycle:
Sets whether or not (YES or NO) you want the
wash cycle to repeat itself at a preset time interval (see
last function step “Cycle Interval”) instead of occurring by
normal scheduling with the “SCHEDULE WASH/CAL”
screen. With “Cycle:” highlighted, use
×
×
or
Ø
Ø
key
to se-
lect the choice.
•
Hold Release:
Sets how long (0-999 seconds) you want
analog Outputs 1 and 2 to remain “held” after the elec-
trode is inserted back into the process. (A cycle begins
with the analyzer automatically “holding” the analog out-
puts. The electrode then retracts. The monitored TTL
inputs determine when the electrode is fully retracted. If
retraction takes longer than 25 seconds, the cycle aborts
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com