PART FOUR - SERVICE AND MAINTENANCE
SECTION 1 - GENERAL INFORMATION
Model P63 pH Analyzer
Rev. 7-1201
100
1.4 Replacing Fuse(s)
The analyzer will have one or two 1/2 amp fuses (250 VAC;
type M or T slow-blow; 5 mm x 20 mm size) depending on
whether it is equipped with the single or dual fusing option.
WARNING:
DISCONNECT LINE POWER TO AVOID THE POSSI-
BILITY OF ELECTRICAL SHOCK.
1.
After disconnecting line power, remove the power sup-
ply board assembly by performing steps 1 through 3 in
the previous “Replacing Relays” Section 1.3.
2.
Remove the blown fuse and replace it with a GLI fuse
or an equivalent (see PART FIVE for fuse part number).
3.
Reinstall the power supply board assembly following
steps 1 through 3 in the previous Section 1.3 in reverse
order.