PART TWO - INSTALLATION
SECTION 3 - ELECTRICAL CONNECTIONS
Model P63 pH Analyzer
Rev. 7-1201
24
Hookup Using Built-in
Temperature Element
☞
Hookup Using External
Pt 1000 ohm RTD
Refer to Figure 2-5 and connect the sensor (or intercon-
nect) cable wires to TB1, matching colors as indicated.
NOTE:
For GLI Differential sensors that have two shield
wires, always connect its outer shield to the green
ground screw (earth ground) and its inner shield to
“SHLD” Terminal 6.
FIGURE 2-5
Connecting GLI Differential Technique Sensor
(with built-in temperature element)
1.
Access the small white jumper on the circuit board
mounted to the back of the analyzer door:
a) Open the analyzer door and remove the ribbon ca-
ble connector from the back of the door.
b) Unfasten the four screws and remove the plastic
cover on the back of the door.
2.
Cut or remove the small white jumper located to the left
of Terminal 10 on TB1.
3.
Refer to Figure 2-6 and connect the sensor (or inter-
connect) cable wires – except the yellow wire – to TB1,
matching colors as indicated.
4.
Tape the end of the unused yellow wire to prevent in-
advertent contact.
Important: Do not cut off this wire
,
which would permanently disable the GLI Differential
sensor’s built-in temperature element.