PART TWO - INSTALLATION
SECTION 3 - ELECTRICAL CONNECTIONS
Model P63 pH Analyzer
Rev. 7-1201
26
3.3 Other Electrodes
3.4 TTL Inputs
3.
Connect a jumper between “REFERENCE” Terminal 3
and “GND” Terminal 7 on TB1.
4.
Connect the electrode’s integral temperature sensor (if
equipped with one) or an external temperature sensor,
to “Pt100/Pt1000 RTD” Terminals 8 and 10 on TB1.
FIGURE 2-7
Connecting Conventional Combination pH Electrode
Alternate 3-wire hookup:
For best temperature accuracy,
especially with long cable lengths, connect the external
temperature sensor to Terminals 8, 9, and 10 on TB1 as
shown in Figure 2-6 and disregard the terminal designa-
tions.
Other types of pH electrodes, including electrodes with tri-
axial cable, electrode pairs, etc. can be used with this
analyzer. To connect one of these types of pH electrodes,
please contact the GLI Customer Service Department for
wiring details.
The analyzer TTL inputs are dedicated for only operating a
GLI Cal-Clean™ System or other automated wash/cal sys-
tem. Refer to PART THREE, Section 5.4 for more details.