5394227, 12
LOGIQ S8/LOGIQ E8 S
ERVICE
M
ANUAL
9 - 32
Section 9-8 - Electrical Safety Tests
Mains on applied part (cont’d)
Where applicable, a typical test setup of non-TEE Probes can be as illustrated in
.
A typical test setup for TEE probes could be as indicated below:
Figure 9-14 TEE Probe Leakage Isolation (Sink) Current Test
The test passes when the reading measure less than the values in: 9-13 on page 9-19.
WARNING
WARNING
The handle of the TEE probes must not be immersed
Содержание LOGIQ S8