GE H
EALTHCARE
D
IRECTION
5394227, 12
LOGIQ S8/LOGIQ E8 S
ERVICE
M
ANUAL
Section 7-5 - Gathering Troubleshooting Data (R4.2.5x and later)
7 - 43
Collecting a screen capture with logs (cont’d)
Trouble image without patient information
To collect a trouble image you need to enable
Transfer Images
(captured without patient information)
to GE under
ADMIN > System Admin
.
To collect the image, press the
P1
key (copy to hard drive). This places an unidentified .jpeg image in
the D:\Log \DL folder. This folder will be transferred to the back office if a transfer is scheduled.
Figure 7-47 Example of a Trouble Image Collected with P1
To collect logs using
Gather Logs
, the image or images will be captured in the log collection Zip file and
will be deleted from the D:\Log \DL folder.
Figure 7-48 Example of a Trouble Image Collected with Gather Logs