Section 7-8 - Error messages
7 - 91
Section 7-8
Error messages
Table 7-20:
Error messages
The system has detected the lower air filter requires cleaning.
Please clean the lower filter.
1. Shutdown the system.
2. Clean the air filter according to "How often should maintenance tasks be performed?"
Cause: CPU Temperature exceeded threshold (
95°C
).
System temperature is too high. System will shut down.
1. Shutdown the system.
2. Clean the air filter according to "How often should maintenance tasks be performed?"
Cause:
CPU Temperature exceeded threshold (100°C) and therefore, system
must be shut down immediately.
System voltage fault. System will shut down.
1. Select
OK
and reboot the system.
2. If the same message appears after reboot, shut down the system and turn off the
breaker. Then turn on the system according to "Power On/Off" on page 3-17.
Cause: Hardware and/or Voltage error detected.
System Error. Please reboot the system.
1. Select
OK
and reboot the system.
2. If the same message appears after reboot, shut down the system and turn off the
breaker. Then turn on the system according to "Power On/Off" on page 3-17.
Cause:
Unrecoverable error occurred and unable to continue scan.
Invalid Hardware Configuration. System will not scan.
1. Select
OK
and reboot the system.
2. If the same message appears after reboot, shut down the system and turn off the
breaker. Then turn on the system according to "Power On/Off" on page 3-17.
Battery deterioration
When the system detects the battery deterioration, the “Battery deterioration” dialog
displays.
Replace the battery for proper function.
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......