GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
5394227, 12
LOGIQ S8/LOGIQ E8 S
ERVICE
M
ANUAL
Section 4-4 - Functional Checks
4 - 47
4-4-23
FibroScan module (Option)
NOTE:
This test is only applied if the FibroScan option has been installed
Perform following tests. Make sure cables are well connected between system and FibroScan Module.
1.) Boot up and wait until the system show scan screen.
2.) Check the 3 LED (green, blue and white) on FibroScan Module front panel are lit.
Figure 4-35 LED on FIbroScan
3.) Select the assigned key to start FibroScan mode - Screen will be dual screen, FibroScan view will
be showed on right side of monitor, and Touch panel will be FibroScan tab.
4.) Connect either M+ or XL+ probe to FibroScan Module.
5.) Select the exam type button to activate the connected probe - The activated probe blue LED will
start blinking.
6.) Apply pressure to probe with you finger tips.
7.) Check on the FibroScan view that probe pressure indicator, A mode and M mode of FibroScan view
is showed.
Figure 4-36 A mode and M Mode of FibroScan
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......