4 - 26
Section 4-4 - Functional Checks
4.) Use the trackball to position the cursor over the required area of the image.
4-4-7-2
Checks
These Image Controls are used to optimize the M mode image. Verify that all the listed controls are
working as intended:
-
Check Horizontal sweep to optimize the display resolution.
-
Check Gain and TGC controls to adjust the range to be imaged.
-
Use the Frequency (move to higher frequencies) or the Frame rate control (move to lower
frame rate) to increase resolution in image.
-
Use the Frequency (move to lower frequency) to increase penetration.
-
Check Focus to move the focal point(s) around the region of interest in the M-Mode display.
-
Check Dynamic range to optimize the useful range of incoming echoes to the available grey
scale.
-
Check Compress and Edge Enhance to further optimize the display.
-
Check Reject to reduce noise while taking care not to eliminate significant low-level diagnostic
information.
Figure 4-21 M-mode Screen Example
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......