Section 4-4 - Functional Checks
Figure 4-28
6.) Connect any probe. (if not connected)
7.) Go to Live CF Mode. (Press “CF” Mode button, unfreeze.)
8.) Verify both B button and CF button are lit GREEN.
9.) Turn “GAIN” encoder around CF Mode button and verify ROI gain change.
10.)Operate TGC sliders and verify B image gain change.
11.)Press “COMMENT” key.
12.)Type in arbitrary strings from A/N Keyboard. Verify keyboard is functional.
4-4-14-2
DVD Drive Test
1.) Insert Blank CD-R or DVD-R Media into Optical Drive.
2.) Select
UTILITY
->
SYSTEM
->
Backup
Restore.
3.) Select “CD/DVD” in “MEDIA” list.
4.) Select “User Define Configuration checkbox” in Backup and Press “Backup”.
5.) Start backup into Media by pressing “OK” Button.
6.) Verify “Finished OK” message appears.
4-4-14-3
LAN Port Test
1.) Plug the Ethernet Cable to connect local area network.
2.) Verify that Local area network connection is recognized.
comment
TGC
CF
B
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......