GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
5394227, 12
LOGIQ S8/LOGIQ E8 S
ERVICE
M
ANUAL
4 - 10
Section 4-3 - General procedures
4-3-6-2
To move the operator panel from side to side
NOTE:
The operator panel can rotate approximately 15 degrees to each side.
Figure 4-7 Swivel Lock Lever (R1 - R3 system)
Figure 4-8 Swivel Lock Lever (R4 system)
1.) Hold Swivel Lock lever under the operator panel.
2.) Rotate the operator panel (while holding lever).
3.) Verify the operator panel is free to rotate.
4.) Release the swivel lock lever (disengage).
5.) Verify the operator panel does not rotate.
NOTICE
!! NOTICE:
Do NOT attempt to apply excessive rotating force to the operator panel. The swivel brake is designed
to hold the operator panel in desired position during normal use, and not meant to completely fix the
the operator panel position.