GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
5394227, 12
LOGIQ S8/LOGIQ E8 S
ERVICE
M
ANUAL
4 - 24
Section 4-4 - Functional Checks
4-4-6-2
Preparations
Use a phantom (optional) when doing these tests.
1.) Connect one of the probes, to the scanner’s active probe connector.
-
see:
3-6-4 "Connecting Probes" on page 3-15
for info about connecting the probes.
-
For available probes, see:
Section 9-13 "Probes" on page 9-59
2.) Turn ON the scanner. The B-Mode window is displayed (default mode).
3.) If needed, adjust the Display’s Brightness and Contrast setting (see:
4-4-6-3
Checks
1.) Press
B-Mode
on the Operator Panel to access B-Mode.
2.) These Image Controls are used to optimize the B-Mode picture:
-
Use
Gain
and
TGC
controls to optimize the overall image together with the
Power
control.
-
Use
Depth
to adjust the range to be imaged.
-
Use
Focus
to center the focal point(s) around the region of interest.
-
Use
Frequency
(move to higher frequencies) or
Frame rate
(move to lower frame rate) to
increase resolution in image.
-
Use
Frequency
(move to lower frequency) to increase penetration.
-
Use the control to optimize imaging in the blood flow regions and make a cleaner, less noisy
image.
-
Use
Reject
controls to reduce noise in the image.
•
Check Width, Focus, Frame rate, Frequency
The results of these adjustments must be verified on the B-Mode sector on the screen.
•
Check Rotation (Up/Down), Reverse (Left/Right), B Color Maps and Cineloop
The results of these adjustments must be verified on the B-Mode sector on the screen.
•
Check Gain, TGC and Depth
•
Check B-Mode Soft Menu Controls
•
Check Compress, Contour, Reject and Tilt
•
Check Acoustic output power and Dynamic Range
WARNING
WARNING
ALWAYS USE THE MINIMUM POWER REQUIRED TO OBTAIN ACCEPTABLE IMAGES
IN ACCORDANCE WITH APPLICABLE GUIDELINES AND POLICIES.
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......