Section 9-6 - System maintenance
9 - 7
Section 9-6
System maintenance
9-6-1
Preliminary checks
The preliminary checks take about 15 minutes to perform. Refer to the system user documentation
whenever necessary.
Table 9-3
System checks
Step
Item
Description
1
Ask & Listen
Ask the customer if they have any problems or questions about the equipment.
2
Paperwork
Fill in the top of the Ultrasound Inspection Certificate (see:
"Ultrasound INSPECTION CERTIFICATE" on
).
Note all probes and system options.
3
Power up
Turn the system power on and verify that all fans and peripherals turn on.
Watch the displays during power up to verify that no warning or error messages are displayed.
4
Probes
Verify that the system properly recognizes all probes.
5
Displays
Verify proper display on the LCD monitor and Touch Panel.
6
Presets
“Full Backup” all customer presets on Hard disk and/or DVD (see:
Section 8-7 "Settings - Backup and
7
Image Archive
Backup the Image Archive on DVD-R, CD-R, USB drives, etc. (see:
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......