Section 6-3 - Volume Navigation Calibration Procedure
Section 6-3
Volume Navigation Calibration Procedure
6-3-1
Purpose of this section
The Volume Navigation Calibration Procedure within the LOGIQ S8 is described in this section.
6-3-2
Overview
The Volume Navigation system is factory calibrated. However, small variations in the electromagnetic
sensors, sensor brackets and the probes themselves can create system-to-system variation. This
procedure is a means to do a system-specific calibration correction.
Note that the field calibration only applies to the specific type of probe (9L-D, ML6-15-D, C1-5-D, e.g.)
that is used. Therefore, it may be necessary to repeat this procedure for each probe type that a
customer owns.
6-3-3
Equipment
Use a phantom (optional) with distinguishable points that can be scanned from at least two directions
(top and side, e.g.). It is best to use a special “calibration phantom” (optional) that has target points in
the phantom (optional) and has been developed for this purpose. Note that the strings in many
phantoms (optional) are not a good choice because it is not easy to specify a particular point along the
string.
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......