Section 4-4 - Functional Checks
4 - 33
4-4-13
Basic Measurements
NOTE:
The following instructions assume that you first scan the patient and then press Freeze.
4-4-13-1
Check Distance and Tissue Depth Measurement
1.) Press
Measure
once to display an active caliper.
2.) Move the Trackball to position the active caliper at the start point (distance) or the most anterior
point (tissue depth).
3.) Press
Set
to fix the start point.
4.) The system fixes the first caliper and displays a second active caliper.
5.) Move the Trackball to position the second active caliper at the end point (distance) or the most
posterior point (tissue depth).
6.) Press
Set
to complete the measurement. The system displays the distance or tissue depth value in
the measurement results window.
NOTE:
Before you complete a measurement:
•
To toggle between active calipers, press
Measure
.
•
To erase the second caliper and the current data measured and start the measurement again, press
Clear
once.
NOTE:
To rotate through and activate previously fixed calipers, turn Cursor Select.
NOTE:
After you complete the measurement, to erase all data that has been measured to this point,
but not data entered onto worksheets, press
Clear
.
4-4-13-2
Check Circumference/Area (Ellipse) Measurement
1.) Press
Measure
once to display an active caliper.
2.) Move the Trackball to position the active caliper.
3.) Press
Set
to fix the start point.
4.) The system fixes the first caliper and displays a second active caliper.
5.) Move the Trackball to position the second caliper.
6.) Turn the Ellipse control; an ellipse with an initial circle shape appears.
NOTE:
Be careful not to press the
Ellipse
control as this activates the Body Pattern.
7.) Move the Trackball to position the ellipse and to size the measured axes (the calipers).
8.) To increase the size, turn the Ellipse control in a clockwise direction. To decrease the size, turn the
Ellipse control in a counterclockwise direction.
9.) To toggle between active calipers, press
Measure
.
10.)Press
Set
to complete the measurement.
11.)The system displays the circumference and area in the measurement results window.
NOTE:
Before you complete a measurement:
•
To erase the ellipse and the current data measured, press
Clear
once. The original caliper is
displayed to restart the measurement.
•
To exit the measurement function without completing the measurement, press
Clear
again.
4-4-13-3
Worksheets
Measurement/Calculation worksheets are available to display and edit measurements and calculations.
There are generic worksheets as well as Application specific worksheets. The worksheets are selected
from the Measurement Touch Panel.
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......