GE H
EALTHCARE
D
IRECTION
5394227, 12
LOGIQ S8/LOGIQ E8 S
ERVICE
M
ANUAL
9 - 8
Section 9-6 - System maintenance
9-6-2
Functional checks
The functional checks take about 60 minutes to perform. Refer to the Basic User Manual whenever
necessary.
9-6-2-1
System functional checks
.
NOTE:
* Some software may be considered standard depending upon system model configuration.
Table 9-4
System Functional Checks
Check
Step
Description
1
B Mode
Verify basic B-Mode (2D) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
2
Color Flow Mode
Verify basic CF-Mode (Color Flow Mode) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
3
Doppler Mode
Where applicable, verify basic Doppler operation (PW and CW if available).
Check the basic Ultrasound system controls that affect this mode of operation.
4
M-Mode
Verify basic M-Mode operation.
Check the basic Ultrasound system controls that affect this mode of operation.
5
3D Mode
Where applicable, verify basic 3D Mode operation.
Check the basic system controls that affect this mode of operation.
6
Real Time 4D Mode
Where applicable, verify basic Real Time 4D Mode operation.
Check the basic system controls that affect this mode of operation.
7
Basic Measurements
Check Distance and Tissue Depth Measurement.
8
Probe Elements
Perform an Element Test on each probe to verify that all the probe elements and Ultrasound
system channels are functional, where applicable.
9
Applicable Software
Options
Verify the basic operation of all optional modes.
Check the basic system controls that affect each options operation.
10
System Diagnostic
Perform the Automatic Tests, where applicable.
11
Transmit/Receive
Verify that all system XMIT/RECV channels are functional, where applicable.
12
Operating Panel test
Perform the Operating Panel Test Procedure.
13
Keyboard
Where applicable, verify basic keyboard functions (for example: type annotation, press F1 to
display on online help.).
14
Touch Panel
Where applicable, verify basic Touch Panel display functions.
15
Monitor
Verify basic monitor display functions
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......