Section 7-3 - Gathering Troubleshooting Data (R1 to R4)
7 - 15
7-3-7 Capturing Network Logs with Network Sniffer (cont’d)
Use filters to limit the captured data. Filter by IP or by IP and port number.
By IP only: press on Capture Filter.
3.) Select
IP only
.
Use the IP address of the system or the one from the DICOM device under test, on the Filter string field,
using the following syntax: (example) host 3.62.12.33 as shown in below, press
OK
.
Figure 7-14 Capture Filter IP Only