FDA Acoustic Output Data
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
6–73
C721 (cont’d)
Imaging
Mode
Maximum
TIC
f
c
W
0
Dimensions of
A
aprt
FL
Focal Zone
Pulsed Wave Doppler
0.20
5.1
8.3
0.9
3.0
B/Pulsed Wave Doppler
0.43
5.1
17.5
0.9
3.0
B/Color Flow Doppler/
Pulsed Wave Doppler
0.46
5.1
19.0
0.9
3.0
Color Flow with M Doppler
0.27
5.2
11.3
0.9
3.0
Table 6–72. C721 Maximum Cranial Bone Thermal Index (TIC)
Imaging
Mode
Scanning
Mode
Maximum
MI
Pr.3
[MPa]
Zsp
[cm]
fc
[MHz]
Pd
[
µ
sec]
PRF
[Hz]
Prl3
@PIImax
B–Mode
Scan
0.46
0.90
3.51
5.46
0.14
8658
0.90
M–Mode
Non–Scan
0.46
0.60
3.51
5.46
0.15
865
0.60
B/M–Mode
Scan
Non–Scan
0.46
0.46
0.90
0.60
3.51
3.51
5.46
5.46
0.14
0.15
8658
865
0.90
0.60
Pulsed Wave Doppler
Non–Scan
0.42
1.00
3.09
5.04
0.53
650
1.00
B/Pulsed Wave
Doppler
Scan
Non-Scan
0.46
0.42
0.90
1.00
3.51
3.09
5.46
5.04
0.14
0.53
8658
650
0.90
1.00
B/Color Flow Mode
Scan
0.46
0.90
3.06
5.04
0.55
280
0.90
B/Color Flow Mode/
Pulsed Wave Doppler
Scan
Non-Scan
0.46
0.42
0.90
1.00
3.06
3.09
5.04
5.04
0.55
0.53
280
650
0.90
1.00
M/M–CFM
Non-Scan
0.46
0.70
3.06
5.04
0.55
280
0.70
M–CFM/PD
Non-Scan
0.46
0.70
3.06
5.04
0.55
280
0.70
CWD Mode
Non-Scan
–
–
–
–
–
–
–
Table 6–73. C721 Maximum Mechanical Index (MI)
Содержание Logiq 500
Страница 1: ......
Страница 30: ...Sub Menu Displays LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 2 12 This page left blank intentionally...
Страница 46: ...3D Surface Mode LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 3 16 This page left blank intentionally...
Страница 376: ...OB Tables LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 7 42 This page left blank intentionally...
Страница 426: ...VCR Operating Instructions LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 8 50 This page left blank intentionally...
Страница 448: ...Worklist Option LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 9 22 This page left blank intentionally...