Advanced Cardiac Calculations
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
4–79
B-Mode Analysis – Parasternal Short Axis (PSAX-PAP)
LVAPd
LVAPs
SYSTOLE
DIASTOLE
Figure 4–67. PSAX–PAP Measurements
Calc Mnemonic
Calc Name
Input Measurements
Formula
LVAPd
Left Ventricular Area,
Papillary Muscles at
Diastole
one area (by ellipse, trace
or circle)
LVAPd=a1[cm^2]
LVAPs
Left Ventricular Area,
Papillary Muscles at
Systole
one area (by ellipse, trace
or circle)
LVAPs=a1[cm^2]
HR
Heart Rate (beats/minute)
one 2 beat time
interval
HR[BPM]=120 [sec]/ 2
beat time [sec]
Table 4–15. Parasternal Short Axis - Papillary Muscles (PSAX-PAP)