Realtime Doppler Calculations
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
4–7
Transferring Calcs
Selecting Transf. CALCs copies data from the Realtime
Doppler Calculation window to the selected location on the
report page.
The following is an example of transferring calculations in the
OB exam category:
1.
Select OB PI, OB RI and OB S/D (D/S) from the
Doppler Realtime Calculation : Submenu selection in
the Set Up/Preset Program menu page 7.
2.
Turn on the Realtime Doppler Calculation selection in
the Softmenu.
3.
The selected measurements will be displayed in the
upper left portion of the screen.
4.
Press Freeze.
5.
Choose “Select Location” from the Measurement
Softmenu.
6.
Select “Transf. CALCS” to transfer the calculations to
the selected location on the report page.
The Transf. CALCs function is not available unless values are
displayed in the Doppler Realtime Display window.
NOTE: The Advanced Vascular Calculation Sub-Menu has an
option called Manual Calc. When Manual Calc is activated,
Auto Calc measurements will not be transferred to the Report
Page or result window.
Содержание Logiq 500
Страница 1: ......
Страница 30: ...Sub Menu Displays LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 2 12 This page left blank intentionally...
Страница 46: ...3D Surface Mode LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 3 16 This page left blank intentionally...
Страница 376: ...OB Tables LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 7 42 This page left blank intentionally...
Страница 426: ...VCR Operating Instructions LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 8 50 This page left blank intentionally...
Страница 448: ...Worklist Option LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 9 22 This page left blank intentionally...