FDA Acoustic Output Data
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
6–49
B510
Mode
MI
TIS/TIB
Scan
TIS
Non-Scan
TIB
Non-Scan
TIC
I
SPTA.3
B-Mode
0.42
0.25
–
–
0.27
8.90
M-Mode
0.42
–
0.02
0.04
0.02
9.60
B/M-Mode
0.42
0.17
0.02
0.04
0.20
15.53
Pulsed Wave Doppler
0.50
–
0.21
0.32
0.24
101.80
B/Pulsed Wave Doppler
0.50
0.17
0.21
0.32
0.41
107.73
B/Color Flow Doppler
0.42
0.28
–
–
0.30
46.93
B/Color Flow Doppler/
Pulsed Wave Doppler
0.50
0.22
0.11
0.16
0.36
77.33
M-Mode/Color M-Mode
0.35
–
0.11
0.11
0.13
19.20
Color Flow with M Doppler
0.35
–
0.11
0.11
0.13
19.20
CW Doppler
0.02
–
0.36
0.67
0.40
147.26
Table 6–31. B510 Maximum Acoustic Output Levels
CWD2
Mode
MI
TIS/TIB
Scan
TIS
Non-Scan
TIB
Non-Scan
TIC
I
SPTA.3
Color Flow with M Doppler
0.098
–
0.49
3.296
1.21
280.805
Table 6–32. CWD2 Maximum Acoustic Output Levels
CWD5
Mode
MI
TIS/TIB
Scan
TIS
Non-Scan
TIB
Non-Scan
TIC
I
SPTA.3
Color Flow with M Doppler
0.06
–
1.919
3.549
3.557
485.1999
Table 6–33. CWD5 Maximum Acoustic Output Levels
Содержание Logiq 500
Страница 1: ......
Страница 30: ...Sub Menu Displays LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 2 12 This page left blank intentionally...
Страница 46: ...3D Surface Mode LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 3 16 This page left blank intentionally...
Страница 376: ...OB Tables LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 7 42 This page left blank intentionally...
Страница 426: ...VCR Operating Instructions LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 8 50 This page left blank intentionally...
Страница 448: ...Worklist Option LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 9 22 This page left blank intentionally...