FDA Acoustic Output Data
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
6–18
FDA Acoustic Output Tables
Maximum output summary
The following tables list the typical maximum acoustic output
levels achievable with the LOGIQ
500 for all probes and
operational modes. It is intended that this information be useful
in making ALARA decisions and selecting the most appropriate
probe for the application. In accordance with US FDA
Guidelines, the overall maximum acoustic SPTA intensity for
LOGIQ
500 is limited to 720 mW/cm
2
and MI is limited to 1.9.
Modes for which TI does not exceed 0.1 are indicated by <0.1.
Содержание Logiq 500
Страница 1: ......
Страница 30: ...Sub Menu Displays LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 2 12 This page left blank intentionally...
Страница 46: ...3D Surface Mode LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 3 16 This page left blank intentionally...
Страница 376: ...OB Tables LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 7 42 This page left blank intentionally...
Страница 426: ...VCR Operating Instructions LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 8 50 This page left blank intentionally...
Страница 448: ...Worklist Option LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 9 22 This page left blank intentionally...