Quality Assurance
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
5–17
Setting up a Record Keeping System
Preparation
The following is needed:
Quality Assurance binder
Magnetic photo pages for hard copy images, or Floppy disk
holder for archived images
Quality Assurance Checklists.
Display the following information while testing quality
assurance:
Acoustic Output
Gain
Depth
Probe
Dynamic Range
Set up new patient to be the name of the test.
Annotate the following:
Any control where its value is NOT displayed.
Significant phantom information.
Record Keeping
Complete the following:
1.
Fill out the Ultrasound Quality Assurance Checklist for
each probe, as scheduled.
2.
Make a hard copy or archive the image.
3.
Compare images to baseline images and acceptable
values.
4.
Evaluate trends over previous test periods.
5.
File hard copy/archive floppy and checklist in Quality
Assurance binder.