Advanced Cardiac Calculations
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
4–66
Advanced Cardiac Specification Tables
The following tables show each first layer menu selection with
the measurements and calculations found in it’s second layer
menu.
The tables show the mnemonic, a description of that mnemonic
(calc name), input measurements required and formula used.
LV Calculation Formulas (Cubed Method)
B-Mode
M-Mode
Figure 4–55. Cubed Method Measurements
Refer to Chapter 10 (Cardiology) of the Basic User Manual for
the following Cubed Method Formulas:
LVIDd, LVIDs, IVSd, IVSs, LVPWd, LVPWs, HR, ET, EdV, EsV,
FS, SV, EF and CO.
Calc Mnemonic
Calc Name
Input Measurements
Formula
SI
Stroke Volume Index
body surface area and SV
SI[ml/m^2]=SV/BSA
CI
Cardiac Index
body surface area and CO
CI[ml/min/m^2]=CO/BSA
MVCF
Mean Vcf
two distances & one time
interval
MVCF[circ/s]=(LVIDd-
LVIDs)/(LVIDd x ET)
LVM
Left Ventricle
Cardiac Mass
three distances
LVM[g]=1.04*[(IVSd+
LVPWd+LVIDd)^3–
LVIDd^3]–13.6
Table 4–2. Cubed Method Calculation Formulas
Содержание Logiq 500
Страница 1: ......
Страница 30: ...Sub Menu Displays LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 2 12 This page left blank intentionally...
Страница 46: ...3D Surface Mode LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 3 16 This page left blank intentionally...
Страница 376: ...OB Tables LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 7 42 This page left blank intentionally...
Страница 426: ...VCR Operating Instructions LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 8 50 This page left blank intentionally...
Страница 448: ...Worklist Option LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 9 22 This page left blank intentionally...