Advanced Cardiac Calculations
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
4–90
Doppler Analysis – Aortic Valve (D-AV)
AOPFV
LOPFV
s
e
P
a
d
F
O
R
V
p
2
Figure 4–77. D-AV Measurements
Calc Mnemonic
Calc Name
Input Measurements
Formula
AOs
Aortic Root
Dimension, Systole
one distance
AOs=d1[cm or mm]
FVI-AV
Flow Velocity
Intergal at Aortic Valve
flow velocities
FVI-AV[mm or cm]=
Vdt
FVI-LV
Flow Velocity Intergal at
Left Ventricular Outflow
Tract
flow velocities
FVI-LV[mm or cm]=
Vdt
PFVAV
Peak Flow Velocity at
Aortic Valve
one velocity
PFVAV=v1[cm/s or m/s]
PFVLV
Peak Flow Velocity at Left
Ventricular Outflow Tract
one velocity
PFVLV=v1[cm/s or m/s]
ET
Ejection Time
one time interval S–E
ET=t1[ms or sec]
PHT
Pressure Half Time
one time interval
PHT=t1[ms orsec]
HR
Heart Rate (beats/minute)
one 2 beat time
interval
HR[BPM]=120 [sec]/ 2
beat time [sec]
Table 4–25. Aortic Valve (D-AV)
Содержание Logiq 500
Страница 1: ......
Страница 30: ...Sub Menu Displays LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 2 12 This page left blank intentionally...
Страница 46: ...3D Surface Mode LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 3 16 This page left blank intentionally...
Страница 376: ...OB Tables LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 7 42 This page left blank intentionally...
Страница 426: ...VCR Operating Instructions LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 8 50 This page left blank intentionally...
Страница 448: ...Worklist Option LOGIQ 500 Advanced Reference Manual 2276614 100 Rev 0 9 22 This page left blank intentionally...