Electrical Requirements
5.6 Typical Customer Supplied Cables and Wiring
5.6.1 Power Input Distribution Box (PIDB) with MDP (A1)
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
5-16
Table 5-6: D670 Customer Supplied Cables and Wiring
Wiring
Description
Maximum run length
Meter (ft.)
Quantity
Size AWG
(mm
2
)
From
To
Run No. 1 – from primary power source to MDP
(see
)
3
*
*
Table 5-3, D670 Minimum Feeder Wire Size, p.5-4
for AWG (mm2) wire sizes
Power
1 (3)
1 (3)
1
1/0 (50)
Ground
1 (3)
1 (3)
Run No. 2 – from MDP to PDU * (see
3
Power
1 (3)
1 (3)
1
1/0 (50)
Ground
1 (3)
1 (3)
1
Neutral (not required)
1 (3)
1 (3)
Run No. 3 – from MDP to System Emergency Off (SEO) (see
)
2
14 (2)
Power
2 (6)
2 (6)
1
14 (2)
Ground
2 (6)
2 (6)
Run No. 4 – CT PDU to warning light control (see
2
14 (2)
Warning light 24 volt control A3J2-1,2,3,4
N/A
N/A
Run No. 5 – CT PDU to scan room door interlock (see
)
2
14 (2)
Scan room door interlock A3J6-1,2
N/A
N/A