Electrical Requirements
5.6 Typical Customer Supplied Cables and Wiring
5.6.1 Power Input Distribution Box (PIDB) with MDP (A1)
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
5-15
If a UPS is ordered, it is strongly recommended that you order UPS from GE (PN B7999PP). In this case, you
must also order the MDP that is designed to work with the UPS.
GE-orderable MDP units (A1) include:
For the U.S. (440V or 460-480V) – PN E4502AB (90A)
For Europe and Asia (380-400V or 420V) – PN E4502AC (110A) (or equivalent).
Table 5-6, D670 Customer Supplied Cables and Wiring, p.5-16
lists the technical details for customer-supplied
wiring.
NOTE
The GE orderable MDP, such as A1, has an Emergency OFF (EMO) switch integrated into the front panel.
Figure 5-4:D670 PIDB Detailed Description
Default terminal for input power
cable from MDP
Alternative location for MDP power
cable terminal
Note:
The power cable terminal can
be moved from the top of the
PIDB (default) to the left or
right side, depending on the
site plan.
Terminals for cables
leading to PDU and
NM gantry