Electrical Requirements
5.5 System Cable Information
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
5-13
Injector
Console
(adaptor box)
Y
8.2
2.5
Power cable injector to console
ECG
CT gantry
N
30
9.1
Gantry to ECG monitor
NM gantry
CT gantry
N
126
29.5
9
CT table interface cable
NM gantry
CT gantry
N
124
29.5
9
NM-CT gantry interface cable
NM gantry
CT gantry
N
125
29.5
9
ECG and respiratory gating cable
NM gantry
Patient table
N
130
65.6 (62.3)
20 (19)
Table cable harness
‡
NM gantry
PIDB
Y
120
65.6 (54.1)
20 (16.5)
NM mains cable
MDP
PIDB
Y
9
2.5
PIDB-to-MDP power feed
NM gantry
NM host
N
128
65.6 (54.1)
20 (16.5)
NM gantry to Host Ethernet cable
NM gantry
NM host
N
127
65.6 (54.1)
20 (16.5)
NM gantry to Host persistence cable
*
Actual length is the entire cable length, connector to connector, including cable run inside each sub-system.
Usable Length is the portion of the cable routed externally between sub-systems.
†
The MDP can be the A1 or other primary power disconnect device
‡
The standard table cable provided is 20 M. There is an optional 10 M cable available.
Table 5-5: D670 Sub-system Inter-connection Cables(Continued)
Start / Destination
H/V
Separation
(Y or N)
Run
#
Actual Length
(Usable Length)
*
Description
From
To
Feet
Meters