Equipment Description and General Construction Requirements
2.3 Room Structural Requirements
2.3.1 Floor Requirements
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
2-29
LEHR collimator
62
137
2 per system/cart
LEGP collimator
55
121
2 per system/cart
ELEGP collimator
62
137
2 per system/cart
MEGP collimator
103
227
2 per system/cart
HEGP collimator
131
288
2 per system/cart
NM UPS
May vary but no
more than 130
May vary but no
more than 287
4 feet
CT UPS (optional)
350
772
4 wheels; Anchored to floor
Xeleris workstation
(insignificant)
Table 2-2: Weight of Components(Continued)
Component
Weight (kg)
Weight
(lb)
Load Distribution
Comments
CAUTION
!
If the system is installed on a floor type thinner than a 166 mm (6.5") concrete floor, the customer shall,
at their expense, provide acceptable anchoring and mounting methods that meet all structural specifications
provided in sections
and
of this manual.
The customer shall ensure that the floor strength in the collimator cart storage area and along the movement
routes for collimator exchange are suitable for the collimator cart load (approx. 250 kg each).