Equipment Description and General Construction Requirements
2.2 Room Size, Layout and Considerations
2.2.2 System Layout Drawings
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
2-19
Figure 2-12:Typical Elongated Layout
Enables installation, operation and service of the system in safe conditions
Enables relative freedom of movement for operator and service, with operator access on both
sides of the system
Includes installation of a table extender
Collimator carts are stored outside the room
Console in operator room is positioned so that operator can see into the bore
Note
:
Dimensions are in cm
EMO