Regulatory Clearances
D.1 Regulatory Clearances
D.1.4 Additional Regulatory Clearance Information
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
D-11
D.1.4 Additional Regulatory Clearance Information
D.1.4.1 Minimum Room Size (Limited Access)
Servicing of the CT System can be safely performed within the regulatory envelopes; however,
sufficient space must be maintained to remove system covers, and replace large system
components. To achieve this clearance for the gantry, clear space must be available to maneuver
the gantry covers mounted on the service dollies. Surface floor raceway cannot be used in the
egress route areas. OSHA ramps are available. The FE lifting the rear or front cover to avoid floor
obstructions is not an EHS-approved service procedure. One service engineer shall be able to
accomplish all service component replace tasks listed without the need for special tools or
equipment, such as a tube change, detector change, and HV tank.