Temperature: 15° to 30°C (59° to 86°F)
Ambient room temperatures above 30°C (86°F) or below 15°C (59°F) will destabilize processing and
replenisher solution temperatures to the detriment of processing performance. Pay special attention to
room temperatures in wintertime and maintain solution temperatures within designated ranges.
•
Relative Humidity: 30% to 70%
Relative humidities above 70% will result in inadequate printer paper transport or improper paper drying,
producing abnormal prints. Maintain the ambient room relative humidity within the range of 30% to 70%
RH at all times.
•
Ventilation
Provide a vent close to the processor section and maintain adequate ventilation.
•
Illumination: 500 lux or lower
Maintain room illumination levels so that no more than 500 lux falls on the scanner and printer upper
surfaces. Should the equipment main body or paper magazine be exposed to direct outside light or strong
reflected light, use curtains or blinds to shield the equipment from such extraneous light.
REFERENCE:
Two 40W fluorescent lamps provide 500 lux of illumination at a distance of 1 meter (3.3 ft.).
NOTE:
Excessively high indoor illumination levels will adversely affect the quality of finished prints.
1.3.1
Main Control Unit FMPC
To prevent damage to the system, abide by the following points.
• Do not install any application software without Fujifilm’s consent.
• Do not modify the data structure or data in the hard disk.
• Do not open the main control unit without the consent of Fujifilm’s service representative.
• Do not add to or modify expansion cards or RAM cards.
• Do not reboot using the [Ctrl]+[Alt]+[Del] keys.