134
FMPC
7
Reinstall the filter by aligning the fastening tab
with the notch, and turn it 90° clockwise.
8
If the “Setting” temperatures differ from the
values displayed, move the cursor to the
“Measured value” and enter the values recorded
in Step 3.
Range : 30.0 to 60.0 (°C)
9
Click the [Enter] button.
10
Repeat Steps 5 through 9 for the P2.
11
Measure the temperature of the PS1/2/3 tank
solution at the PS2 processing tank and record
it.
12
Repeat Steps 8 and 9 for PS1/2/3 tank solution.
13
Remove the water supply port inner cover and
PS4 sub-tank rubber plug.
14
Measure the temperature of the PS4 tank
solution and record it.
15
Repeat Steps 8 and 9 for the PS4 tank solution.
16
Wash the circulation filter tray.
17
Reinstall the rubber plug, water supply port inner
cover and circulation filter tray.
18
Click the [Close] button.
• Operation returns to the menu screen.
P1 Filter Rod
Tab
Notch
L
2
234
PS2 Processing Tank
Probe
Thermometer
L
2
1048
Probe
Thermometer
L
2
1049