117
FMPC
If, for instance, the calendar timer is set to 5:00
AM and night-time preheating is turned ON at
3:30 AM, all the power will be shut down two
hours later at 5:30 AM.
NOTE:
Night-time preheating is usually conducted when
the room temperature drops below 10 °C (50 °F) at
night and the temperature is not controlled for 8 or
more hours.
(1) Select “YES” for “Preheat Setting”.
• The 1st and 2nd preheat setting boxes appear.
(2) Move the cursor to “1st Preheat” and enter the
time.
(3) Move the cursor to “2nd Preheat” and enter the
time.
NOTE:
For preheating only once, set the same time for
both the first and second times.
• Select “NO” when no preheat timer is used.
• The second preheating timer is ignored if the
first and second times overlap.
<Example>
1st. preheat time:
2:00 AM
2nd. preheat time: 3:00 AM
With the above settings, night-time preheating
will occur from 2:00 AM to 4:00 AM (2 hours).
4
Click the [OK] button.
• Operation returns to the menu screen.
9.2.6
Error Information Check (415)
[Selection]
“4 Setup and Maintenance” - “1 System Operation
Setup and Check” - “5 Error Information Check”
[Procedure]
The system error log lists up to 300 errors in
descending order of occurrence.
1
Click the [Log Info.] button.
2
Click the [OK] button.
• The latest 90 accumulated errors are displayed.