202B1263075
Problem
Cause
Remedy
The endoscopic
image is discolored
during an
examination or
treatment.
1) The imaging section is damaged.
2) The system is malfunctioning possibly due
to static charge.
3) The video signal cable has shorted out or
broken.
1) 2) 3)
If the problem occurred during
treatment
[Note 1]
, stop treatment and remove
the endotherapy device from the endoscope.
Then, reset
[Note]
the video processor and
the light source. If the problem occurred
during an examination, reset
[Note]
the
video processor and the light source. If
an appropriate image does not appear
even after resetting and it is impossible to
continue the examination, turn them off,
straighten the bending section to unlock,
release the angle knobs, withdraw the
endoscope slowly from the patient, and then
contact your local FUJIFILM dealer.
[Note] Reset: Turn off the video processor and the light source, and wait for at least 5 seconds. Turn on the video
processor and the light source again, and then light the lamp by pressing the Lamp button.
[Note 1] During treatment with a snare, withdraw the snare after loosening it enough.
Chapter 5 Troubleshooting
97