3.1.1 Cleaning and Disinfecting (or Sterilizing) the Forceps Valve
The forceps valve is provided without sterilization and must be cleaned and disinfected (or
sterilized) prior to use following the instructions of the Reprocessing Manual.
The forceps valve is a single-use product. Do not reuse it to avoid infection.
[Note] For information on how to clean and disinfect (or sterilize) the forceps valve, refer
to the Reprocessing Manual.
Reprocessing Manual “7.1 Cleaning and Disinfecting (or Sterilizing) Forceps Valve”
3.1.2 Inspecting the Forceps Valve
The forceps valve is intended for single use. If any abnormality is found, do not use the
product, and use a new disinfected (or sterilized) forceps valve.
(1) Visually check that the slit in the cap (a) and hole in
the main body (b) of the forceps valve are free from
abnormalities such as tears, cracks, discoloration, etc.
(2) Attach the lid to the main body of the forceps valve.
3.1.3 Attaching the Forceps Valve
(1) Attach the forceps valve to the instrument channel inlet of
the endoscope.
a
b
Forceps valve (FOV-DV7)
Chapter 3 Preparation and Inspection Before Use
47