202B1263075
(3)
Attach a fixing rubber to the tip of the closed claw section.
[Note]
Deeply insert the claw section into a fixing rubber until the
tip of the claw section protrudes from the fixing rubber.
(4) Slowly pull the trigger until the end surface of the piston
makes contact with the fixing rubber. At this position, the
fixing rubber is kept expanded.
[Note]
Pull the trigger while putting your finger on the tip of the
claw section so as not to disengage the fixing rubber.
[Note]
Do not insert your finger into the fixing rubber. Your
finger may be caught in the fixing rubber.
(5) Insert the distal end of endoscope into the setting tool,
and then align the position of the fixing rubber with the
balloon fixing portion (shaded portion) on the side of the
distal end of endoscope.
(6)
Pull the trigger further. The fixing rubber is pushed out
and is attached to the balloon fixing portion.
[Note]
If the attached fixing rubber becomes twisted, remove the
twists with your finger to attach the fixing rubber.
[Note]
Attach a fixing rubber so as not to cover the balloon air
supply/suction port of the endoscope.
Chapter 3 Preparation and Inspection Before Use
56
EI580BT_E2-10_202B1263075.indb 56