202B1263075
3.4 Connecting Balloon Controller
WARNING
The filter and tube of the tube kit are consumables. Replace them once
a month or once every 10 cases, whichever comes first. If body fluid
flows back into the tube, replace the tube kit. Not doing so may pose
an infection risk.
CAUTION
Connect the endoscope and the Over-tube to the balloon controller
properly. Before insertion, make sure that the balloon operates as
intended. Otherwise, it may cause an injury.
[Note] This endoscope cannot be connected to the balloon controller PB-10.
3.4.1 Connecting the Balloon Controller
(1) Install the balloon controller referring to the operation
manual of the balloon controller.
Connect the tube kit as follows.
Balloon controller
Applicable tube
Tube connector 2
(endoscope side)
TY-06S tube kit provided with
this product or the TY-06 tube kit
(optional item)
Tube connector 1
(Over-tube side)
Tube 1 (white tube) of TY-04 tube
kit or the TY-06 tube kit (optional
item)
[Note]
The filter and tube of the tube kit are consumables.
Replace them once a month or once every 10 cases,
whichever comes first.