202B1263075
(2)
Prepare the fixing rubber setting tool and balloon setting
tool for the balloon to be used.
Table 3.3 Compatible setting tools with the balloons
Balloon
Fixing rubber
setting tool
Balloon
setting tool
BS-2
ST-01B
ST-10 (optional item)
ST-05B
BS-4
ST-10 (optional item) ST-05B
(3)
For BS-2, clean and disinfect the balloon and fixing rubber
according to the Reprocessing Manual.
Reprocessing Manual
(4)
Perform autoclave sterilization for the fixing rubber setting
tool and balloon setting tool according to the operation
manual of the autoclaving unit.
[Note]
The ST-01B fixing rubber setting tool (or ST-10) and
ST-05B balloon setting tool have not been sterilized before
shipment. Be sure to perform autoclave sterilization for
them according to the operation manual of the autoclaving
unit before an examination or treatment.
(5)
Check that the exterior of the fixing rubber setting tool and
balloon setting tool is free from defects such as damage
and folding.
(6) Check the Over-tube according to its operation manual.
Chapter 3 Preparation and Inspection Before Use
51