3.6 Connecting Endoscope (Attachment)
WARNING
Firmly connect the scope connector of the endoscope and the
light source. If the scope connector is not connected properly, the
endoscopic image may flicker or be lost.
CAUTION
Do not attach the video connector of the endoscope to the electrical
connector socket of the VP-7000 with the endoscope inserted into the
scope connector of the BL-7000.
Firmly connect the scope connector of the endoscope and the light
source. Do not look into the connecting part between the endoscope
and the light source. Light leaking from the connecting part may cause
damage to the eyes.
Immediately after removing the LG connector from the light source, do
not touch the light guide rod with hands since it is extremely hot. There
is a risk of burn injury.
Set the suction pressure between 40 and 53 kPa. If the suction
pressure is too high, the endoscope may adhere to mucous membrane,
resulting in damage to the mucous membrane.
3.6.1 Connecting the Endoscope (Video Processor and Light Source)
<When VP-4450HD is used>
(1) Before attaching the LG connector to the light source,
visually check that there is no dirt on the light incident
distal face and no cracks on the glass of the light incident
distal face.
Light incident distal face
Chapter 3 Preparation and Inspection Before Use