202B1263075
(2) Securely insert the endoscope-side connector of the
tube kit (TY-06S) into the balloon air feed inlet of the
endoscope.
(3) Connect the tube 1 (white tube) of the balloon controller
to the air feed inlet (clear tube) of the Over-tube.
3.4.2 Inspecting the Balloon (Balloon Controller)
(1) Make sure that the “ENDOSCOPE” side connector of the
balloon controller is connected to the balloon air feed inlet
of the endoscope, and the “OVERTUBE” side connector
of the balloon controller is connected to the air feed inlet
of the Over-tube.
(2) Press the endoscope-side switch on the balloon controller
to make sure that the balloon attached to the distal end of
endoscope inflates.
(3) Press the switch again to make sure that the balloon
deflates.
Chapter 3 Preparation and Inspection Before Use
61
EI580BT_E2-10_202B1263075.indb 61
2015/08/12 16:20:39