202B1263075
CAUTION
Do not give a strong impact to the LG connector. Install the light source
away from obstacles to prevent the LG connector connected to the
light source from accidental impact damage. During the operation of
an electric bed, etc., ensure that the LG connector connected to the
light source does not hit the bed. Otherwise, the endoscope and light
source may be damaged.
Do not bend the insertion portion near the control portion in a small
radius. It may cause the endoscope to malfunction.
When inserting or withdrawing the endoscope, make sure that the
balloon is deflated. If not deflated, the balloon may burst.
When removing a balloon, take care not to apply excessive force to the
endoscope. It may damage the endoscope.
[Note] Monitor the status of the balloon with X-ray fluoroscopy.
[Note] Only use the balloon controller designated by FUJIFILM.
Chapter 4 Method of Use
75