Problem
Cause
Remedy
Halation
1) The imaging section is damaged.
2) The brightness level is set around “MAX.”
3) The iris mode is set to “AVE.”
1) If the problem occurred during
treatment
[Note 1]
, stop treatment and
remove the endotherapy device from
the endoscope. Then, reset
[Note]
the video
processor and the light source. If the
problem occurred during an examination,
reset
[Note]
the video processor and the light
source. If an appropriate image does not
appear even after resetting, turn them off,
straighten the bending section to unlock,
release the angle knobs, withdraw the
endoscope slowly from the patient, and
then contact your local FUJIFILM dealer.
2) Set the brightness level around 0.
Operation manual of a light source
3) Set iris mode to “PEAK.”
Operation manual of a light source
[Note] Reset: Turn off the video processor and the light source, and wait for at least 5 seconds. Turn on the video
processor and the light source again, and then light the lamp by pressing the Lamp button.
[Note 1] During treatment with a snare, withdraw the snare after loosening it enough.